Model Number 120003 |
Device Problem
Failure to Cycle (1142)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the vent will intermittently not cycle while connected to oxygen.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation was completed.No product was returned for investigation.The cause of the reported problem could not be determined.The reported fault suggests the 15 year old device did malfunction.Without the device being returned for analysis, it is not possible to determine whether this was due to wear and tear or impact damage.Intermittent cycling was also reported in july 2021 and preventive measures were taken so it is unlikely that the reported problem is due to the previous repair.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available.
|
|