• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL PNEUPAC VENTILATORS PARAPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL PNEUPAC VENTILATORS PARAPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number 120003
Device Problem Failure to Cycle (1142)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the vent will intermittently not cycle while connected to oxygen.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation was completed.No product was returned for investigation.The cause of the reported problem could not be determined.The reported fault suggests the 15 year old device did malfunction.Without the device being returned for analysis, it is not possible to determine whether this was due to wear and tear or impact damage.Intermittent cycling was also reported in july 2021 and preventive measures were taken so it is unlikely that the reported problem is due to the previous repair.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS PARAPAC
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
attn:receiving
minneapolis, MN 55442
MDR Report Key15742713
MDR Text Key306811090
Report Number3012307300-2022-26813
Device Sequence Number1
Product Code BTL
UDI-Device Identifier10610586045486
UDI-Public10610586045486
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number120003
Device Catalogue Number120003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-