• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. KIDS KIT CLOSED BLOOD SAMPLING SYSTEM; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. KIDS KIT CLOSED BLOOD SAMPLING SYSTEM; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR Back to Search Results
Catalog Number MX4137
Device Problem Material Separation (1562)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 09/30/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that a kids kit attached to an umbilical arterial line broke off.The infant started bleeding back via the umbilical arterial catheter.A nurse at the bedside noticed the bleeding and promptly clamped the line.
 
Manufacturer Narrative
Other, other text: no product or photographic evidence were provided to aid in this investigation.The complaint report offered insufficient details to determine whether this product functioned as intended, or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.A manufacturing device history record review was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIDS KIT CLOSED BLOOD SAMPLING SYSTEM
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer (Section G)
NULL
201 west queen st.
southington CT 06489
Manufacturer Contact
jim vegel
6000 nathan lane north
3660 s. national avenue
minneapolis, MN 55442
MDR Report Key15744031
MDR Text Key303161210
Report Number3012307300-2022-26826
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMX4137
Device Lot Number4244391
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-