The disposable perforator (id 261221) was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - visual inspection utilizing unaided eye performed: unit was autoclaved, deforming the blue sleeve, and destroying the label, the lot # could not be determined via visual inspection.Spring test attempted: unit had difficulty rotating the hudson end, likely from being autoclaved, but the spring functioned as designed.Function test performed: unit successfully drilled 5 holes with no issues and was found to function as intended.Complaint could not be verified - autoclaving the unit made root cause determination difficult.Unit was found to meet all acceptance criteria.The complaint condition could not be confirmed.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
|