• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED CARELINK WEB UPLOADER; CONTINUOUS GLUCOSE MONITOR RETROSPECTIVE DATA ANALYSIS SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED CARELINK WEB UPLOADER; CONTINUOUS GLUCOSE MONITOR RETROSPECTIVE DATA ANALYSIS SOFTWARE Back to Search Results
Model Number ACC-7350
Device Problems Computer Software Problem (1112); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information received by medtronic indicated that the customer was getting an error message with the carelink pro and stated that the error message was regarding the security check of the carelink uploader application that failed.It was also stated that the customer was having issues with adding the exceptions to the av software.No harm requiring medical intervention was reported.The device will not be returned for analysis.
 
Manufacturer Narrative
Investigation/testing summary: based on error message, it was determined that some form of anti-virus or security software was interfering with uploader application.Met w/ customer on a call and used process monitor application to collect detailed logs about what customer's software was doing when trying to use uploader application.Was able to identify names of processes in logs that indicated that cisco security software was active while customer was using uploader.(most likely) root cause: cisco security software installed on customer's system was attempting to prevent the uploader launch.Analysis summary: determined source of software interference based on error message customer was receiving and collection of detailed diagnostic logs.Provided customer with information as well as guidance on how to add the necessary exclusions and exceptions to their cisco security software.Customer reported issue as resolved after applying recommended fixes.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK WEB UPLOADER
Type of Device
CONTINUOUS GLUCOSE MONITOR RETROSPECTIVE DATA ANALYSIS SOFTWARE
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325
Manufacturer Contact
tricha miles
18000 devonshire st.
northridge, CA 91325
7635140379
MDR Report Key15751520
MDR Text Key307071278
Report Number2032227-2022-356977
Device Sequence Number1
Product Code PHV
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberACC-7350
Device Catalogue NumberACC-7350
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/22/2022
Initial Date FDA Received11/08/2022
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
-
-