• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XPT SARS-COV-2 IGG (SCOVG); SARS-COV-2 IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XPT SARS-COV-2 IGG (SCOVG); SARS-COV-2 IMMUNOASSAY Back to Search Results
Model Number N/A
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2022
Event Type  malfunction  
Manufacturer Narrative
A united states customer contacted siemens customer care center to report the advia centaur xpt sars-cov-2 igg (scovg) lot 015 was unable to be calibrated due to failure of the low calibration deviation.A siemens customer service engineer (cse) was sent to the customer site for system inspection and no issues were identified.The customer was using scovg lot 014 without any calibration issues.Although there is no potential for serious injury in this case, an mdr will be reported to the fda as a requirement of the emergency use authorization (eua).Siemens healthcare diagnostics is investigating.
 
Event Description
The customer was unable to calibrate the advia centaur xpt sars-cov-2 igg (scovg) lot 015 due to failure of the low calibration deviation.The customer observed the expected rlu (relative light units) for the scovg low calibrator as a negative rlu.There are no known reports of patient intervention or adverse health consequences due to the scovg calibration failure.
 
Manufacturer Narrative
Siemens filed the initial mdr 1219913-2022-00386 on november 08, 2022.December 13, 2022 additional information: siemens healthcare diagnostics inc.Investigation confirmed that the failed calibrations are due to the standard s01 value in the master curve card (mcc) for the advia centaur systems lots 19529015 and 27236015, being set at 0.1 index.The s01 value should be 0.01 index.The issue is isolated to the advia centaur systems scovg lots 19529015 and 27236015, smns 1120376 and 1120379.Customers were notified of the issue and instructed to discontinue use of the kits.(us: cc 23-01.A.Us was sent to us customers on 2022-12-16 and outside the us: cc 23-01.A.Ous was sent to customers on 2022-12-16 who have received the advia centaur systems sars-cov-2 igg (scovg) lots 19529015 and 27236015).In section h6, the type of investigation, investigation finding, and investigation conclusion codes were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XPT SARS-COV-2 IGG (SCOVG)
Type of Device
SARS-COV-2 IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avemue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
333 coney street
e. walpole MA 02032
Manufacturer Contact
eiman sulieman
333 coney street
e. walpole, MA 02032
7372808688
MDR Report Key15752164
MDR Text Key307294588
Report Number1219913-2022-00386
Device Sequence Number1
Product Code QKO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EUA202670
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/21/2023
Device Model NumberN/A
Device Catalogue Number11207376
Device Lot Number015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/21/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-