Model Number 0600-0783 |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A replacement glidescope core smart cable was provided to the customer, and the reported smart cable was returned to verathon for evaluation.A verathon technical service representative evaluated the reported smart cable but was unable to confirm the reported issue.When connecting the reported smart cable to known, good, test verathon equipment, the image feed from the cable was normal with no issues found.The camera image quality test was performed and passed.Upon completion of the evaluation the reported glidescope core smart cable was scrapped as it had already been replaced prior to its return to verathon.No further investigation is required at this time.Verathon will continue to monitor for trends.
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope core smart cable, the image intermittently disappears.The procedure was completed using a backup glidescope system, which was made available in about five (5) minutes.No harm to the patient was reported.
|
|
Search Alerts/Recalls
|
|
|