A customer reported that during a patient procedure, using a glidescope bflex 5.0 single-use bronchoscope, when pulling the bronchoscope out of the double lumen tube (41 fr shiley), the camera tip had detached and the distal tip of the bronchoscope was frayed.The physician clamped the tracheal side of the double lumen tube to ensure the forced positive air didn't push anything into the patient's right lung.They worked to safely oxygenate the patient through the bronchial side of the tube.The patient was extubated and reintubated with a 39 fr dlt and a different manufacturer scope.The physician performed a diagnostic view of the right lung bronchi with no reported injury or foreign body.A forceps was used to remove the lodged camera from the distal tip of the 41 fr dlt.X-rays were taken and confirmed no foreign bodies or injury to the patient.
|
The reported glidescope bflex 5.0 single-use bronchoscope was returned to verathon for evaluation.A verathon technical service representative evaluated the returned device and confirmed the reported damage and detached camera.It was determined that the likely root cause of the reported issue is due to inserting the glidescope bflex 5.0 single-use bronchoscope into the double lumen tube used during the procedure.The glidescope bflex bronchoscope - endotracheal tube compatibility" table found in the glidescope bflex single-use bronchoscopes operations & maintenance manual (omm) currently only lists the compatibility of the glidescope bflex single-use bronchoscopes with certain endotracheal tubes, double-lumen tubes, and airway catheters.The double-lumen tube used during the procedure has not been validated by verathon for compatibility for use with the glidescope bflex 5.0 single-use bronchoscope.Any usage of the glidescope bflex bronchoscope outside of what is listed in the omm is outside of verathon's released labeling.Upon completion of the evaluation, the reported glidescope bflex 5.0 single-use bronchoscope was scrapped as it is a single-use device and there being no repairs available for this device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|