SOLTA MEDICAL, INC THERMAGE CPT SYSTEM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACC. Back to Search Results
Model Number TTNS3.00E4-1200
Device Problems Dent in Material (2526); Arcing (2583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2022
Event Type  malfunction  
Event Description
A user facility reported a dented thermage treatment tip.The cpt thermage treatment tip was evaluated and dielectric breakdown was observed.
 
Manufacturer Narrative
The tip passed the flow and thermistor tests but failed the leak test.The tip failed visual inspection as there was both a dent in the middle of tip surface as well as dielectric breakdown observed.No functional testing was performed due to dielectric breakdown.Service confirmed dielectric membrane breakdown on the tip.Breakdown of the membrane can cause the radiofrequency energy, delivered by the system, to focus in a small area of the membrane, rather than to be uniformly distributed over the entire membrane area and could possibly cause patient burns.A review of manufacturing records showed all requirements were met.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Final test verification specifications are acceptable.No non-conformities or anomalies found related to this complaint when reviewing the device history record.It is unknown how the dielectric breakdown occurred as all treatment tips are visually inspected prior to packaging.No corrective action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMAGE CPT SYSTEM TIP
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACC.
Manufacturer (Section D)
SOLTA MEDICAL, INC
11720 n creek pkwy n
ste 100
bothell WA 98011
Manufacturer (Section G)
SOLTA MEDICAL, INC.
11720 n creek pkwy n
ste 100
bothell WA 98011
Manufacturer Contact
juli moore
3365 tree court ind blvd
st. louis, WA 63122
6362263220
MDR Report Key15756995
MDR Text Key307209308
Report Number3011423170-2022-00140
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K132431
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberTTNS3.00E4-1200
Device Catalogue NumberTTNS3.00E4-1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-