Model Number DSX500H11C
Device Problems Degraded (1153); Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer previously reported an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.There was no report of patient harm or injury.The patient also alleged visulization of particles.The device was returned to the manufacturer's quality product investigation laboratory for investigation.The manufacturer visually inspected the external part of the device and found minimal signs of dust contamination in the humidifier chamber.The manufacturer visually inspected the device internally and found minimal dust contamination in the blower assembly.The humidifier showed a small amount of liquid ingress under the bottom plate.The device's event logs were downloaded and reviewed by manufacturer.The manufacturer found no error codes.The manufacturer applied power to the device and verified airflow.The manufacturer concludes there was no evidence of sound abatement foam degradation, but a small amount of liquid ingress and dust contamination were observed and are consistent with being from an external source.Section h6 were updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key15759785
MDR Text Key305376598
Report Number2518422-2022-92731
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500H11C
Device Catalogue NumberDSX500H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received12/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-