Model Number CDHFA500Q |
Device Problems
Wireless Communication Problem (3283); Interrogation Problem (4017)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Event Description
|
It was reported the device was unable to connect to the remote monitoring application via bluetooth (ble).Technical support was contacted and recommended an in clinic follow up for further investigation.No intervention has been performed at this time.The patient was stable and will continue to be monitored.
|
|
Event Description
|
Additional information was received indicating the patient presented in clinic, upon interrogation the device was noted to be in bluetooth (ble) telemetry lockout.Interrogation of the device resolved the event.The patient was stable.
|
|