• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP TEFLON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP TEFLON Back to Search Results
Model Number 3003-009
Device Problem Device Slipped (1584)
Patient Problem Laceration(s) (1946)
Event Date 10/06/2022
Event Type  Injury  
Manufacturer Narrative
As the device did not show any deviation that could cause the reported incident we suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
Customer informed our service department on the (b)(6) that one of our products was involved in a case where a minimal slippage with slight laceration occured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® QR3 SKULL CLAMP TEFLON
Type of Device
DORO® QR3 SKULL CLAMP TEFLON
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
christopher schmitz
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key15768758
MDR Text Key303394584
Report Number3003923584-2022-00016
Device Sequence Number1
Product Code HBL
UDI-Device Identifier04250435500781
UDI-Public04250435500781
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number3003-009
Device Catalogue Number3003-009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/17/2022
Initial Date FDA Received11/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-