• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE Back to Search Results
Model Number M00562301
Device Problems Failure to Deliver Energy (1211); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a captivator small oval stiff snare used to remove multiple gastric polyps in the gastric body during endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure and inside the patient, the physician found that the mucosa of the gastric body was slightly congested with blood through painless gastroscopy and multiple polyp-like bulges were seen, which were removed with a captivator single-use polypectomy snare and cold extraction.During the treatment, the snare was put into the patient's body through the endoscopic channel for polyp removal.During the removal process, it was found that the device had no conductive effect and could not play the role of polyp removal.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.
 
Manufacturer Narrative
The complainant was unable to provide the suspected device lot number; therefore, the lot expiration and device manufacture dates are unknown.The reported (b)(6).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15770406
MDR Text Key307509557
Report Number3005099803-2022-06533
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729019312
UDI-Public08714729019312
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00562301
Device Catalogue Number6230
Was Device Available for Evaluation? No
Date Manufacturer Received10/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-