• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRHYTHM TECHNOLOGIES, INC ZIO AT; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRHYTHM TECHNOLOGIES, INC ZIO AT; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Device Problem Data Problem (3196)
Patient Problem Atrial Fibrillation (1729)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
The patient experienced symptomatic rapid af that was not detected by the device¿s automatic detection feature and was subsequently not transmitted during the wear-period.
 
Manufacturer Narrative
An automated ticket for a suspected leads-off condition was generated and upon follow-up the customer communicated the device had been removed because they had a pacemaker placed and had just been released from the hospital following pacemaker placement surgery.The device was returned to irhythm for processing and an additional finding of rapid af was identified while compiling the final report which was not transmitted during the wear-period.The patient¿s physician, however, had recommended she visit the er based upon a rapid af notification due to a second event, which was transmitted to the account as designed shortly after the missed event.An analysis of the diagnostic data was completed and attributed the missed episode to bluetooth communication issues between the patch and gateway.The zio at ecg monitoring system is intended to capture, analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram (ecg) information for long-term monitoring.While continuously recording patient ecg, both patient-triggered and automatically detected arrhythmia events are transmitted to a monitoring center for reporting.After wear, a final report is generated based on beat-to-beat information from the entire ecg recording.It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety.The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience.It is not intended for use on critical care patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIO AT
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
IRHYTHM TECHNOLOGIES, INC
699 8th st suite 600
san francisco CA 94103
Manufacturer (Section G)
IRHYTHM TECHNOLOGIES, INC
6550 katella avenue, suite 200
cypress CA 90630
Manufacturer Contact
mazi kiani
699 8th st suite 600, CA 94103
9494132147
MDR Report Key15771299
MDR Text Key307775908
Report Number3007208829-2022-00051
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/18/2023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2022
Initial Date Manufacturer Received 10/24/2022
Initial Date FDA Received11/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age73 YR
Patient SexFemale
-
-