The product investigation was completed.Device evaluation details: visual analysis of the returned sample revealed reddish material in the pebax.An electrical test was performed, and no electrical issues were found.Scanning electrode microscope analysis shows evidence of mechanical damage and a hole on the pebax surface.Blood inside the pebax could be related to the noise reported by the customer.A manufacturing record evaluation was performed for the finished device 30814163l number, and no internal action was found during the review.The electrical issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain the following recommendation: ecg noise is typically generated as the result of the improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify the proper connection.It is recommended to turn off the notch filter for this verification.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: pc-(b)(4).
|