Model Number PM2240
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2022
Event Type  Injury  
Event Description
It was reported that the pacemaker exhibited inability to interrogate and loss of pacing.The pacemaker was explanted and replaced.Patient was stable.
 
Manufacturer Narrative
The reported events of inability to interrogate and inability to pace were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The device was cut open to enable further testing and the battery was found depleted.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested by connecting to an external power source.Test results indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the assurity and endurity pacemakers header anomaly advisory issued by abbott on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15774243
MDR Text Key303538532
Report Number2017865-2022-44963
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Model NumberPM2240
Device Catalogue NumberPM2240
Device Lot NumberA000036144
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2022
Initial Date FDA Received11/10/2022
Supplement Dates Manufacturer Received02/20/2023
Supplement Dates FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1530-2021
Patient Sequence Number1
Treatment
RIGHT ATRIAL LEAD; RIGHT VENTRICLE LEAD
Patient Outcome(s) Required Intervention;
Patient Age83 YR
Patient SexMale
-
-