MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY HARMONY LC SURGICAL LIGHTING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY HARMONY LC SURGICAL LIGHTING SYSTEM Back to Search Results
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2022
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite and inspected the harmony dual lc surgical lighting system and found that the lighthead control board and the board to lamp wire harness were damaged resulting in the reported event.The technician replaced the damaged components, tested the unit, confirmed it to be operating according to specification and returned it to service.No additional issues have been reported.
 
Event Description
The user facility reported that during a patient procedure one of the lightheads went out and a burning smell was emitting from their harmony dual lc surgical lighting system.The procedure was completed successfully using another functioning lighthead.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONY LC SURGICAL LIGHTING SYSTEM
Type of Device
LIGHTING SYSTEM
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15777332
MDR Text Key304413188
Report Number1043572-2022-00082
Device Sequence Number1
Product Code FTD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-