Catalog Number MB-35-2X4-8
Device Problems Difficult to Fold, Unfold or Collapse (1254); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in a clear plastic bag.A lot number was not provided with the returned device.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with a clear liquid inside of the catheter.The basket was fully retracted into the tubing and the handle was in the advanced position.During a functional test there was a grinding/scraping sound when the handle was manipulated with no movement from the basket.The handle was disassembled, and it was noted that the drive wire was separated from the handle with nesting of the wire inside the purple hub.For further evaluation of the drive wire cable and basket, the catheter was cut to push the drive wire cable out of the sheath.The basket was fully formed and intact with a dark green substance on the basket wires.Evidence of solder was present on the handle cannula at the joint.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation confirmed the report of unable to open the basket.This report was due to a separation of the device in the handle.The cause of the separation is unknown.The instructions for use also states: "confirm desired position of basket sheath relative to target.Advance basket out of sheath.Caution: pulling on sheath while advancing or retracting basket may damage device, rendering it inoperable." basket deployment difficulties and buckling/breaking of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During a cbd stone extraction, the physician used a cook memory ii double lumen extraction basket.It was reported that when the user attempted to open the basket to extract the stones the basket would not open at the target site.There was no reportable information at this time.The device was received (b)(6)2022.It was noted that the drive wire has disconnected from the handle [subject of this report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY II DOUBLE LUMEN EXTRACTION BASKET
Type of Device
LQR DISLODGER, STONE, BILIARY
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key15779663
MDR Text Key307014518
Report Number1037905-2022-00626
Device Sequence Number1
Product Code LQR
Combination Product (y/n)N
PMA/PMN Number
K171969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMB-35-2X4-8
Device Lot NumberW4599987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2022
Initial Date FDA Received11/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS JF SCOPE
-
-