Concomitant product: e3783r-28asp, lap wire l-hook elect retract (lot#: unknown) device evaluated by mfr evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the returned device have been subjected to improper cleaning or sanitation methods.It was reported that the use of the device resulted in a burn and the device discharged a visible spark or arc.The insulation of the device was damaged, and the device melted.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: tissue buildup (eschar) on the tip of an active electrode may create embers that pose a fire hazard, especially in o xygen-enriched environments.Keep the electrode clean and free of debris.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
According to the reporter, intra operatively, the device outer cylinder on the tip side was melted and sparks appeared from the melted hole during surgery.At first, sparks appeared 5 minutes after using the hook.The electrode was immediately replaced with a spatula-type one, however, a spark also appeared due to proctal prolapse, so it was replaced with a new spatula-type and used.There was also a insulator problem total of 3 samples submitted.There was non serious injury.
|