Model Number CD3357-40C
Device Problem Intermittent Capture (1080)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2022
Event Type  Injury  
Event Description
It was reported that a patient presented with loss of capture noted on the patient's implantable cardioverter defibrillator (icd), causing non-sustained over-sensing episodes.Corrective programming changes were recommended.No patient consequences were reported.
 
Manufacturer Narrative
Further information was requested, but not received.
 
Event Description
It was reported that the device was explanted and replaced on (b)(6) 2023.No patient consequences were reported.
 
Manufacturer Narrative
The reported event of capture anomaly was confirmed.However, the cause of capture anomaly was due to the programmed settings of the device.The device functioned as expected and according to its programmed settings.Visual inspection of the septum, setscrew and contact spring did not reveal any anomaly that could contribute to the reported event.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15784641
MDR Text Key306736086
Report Number2017865-2022-45093
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000095164
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; QUICKFLEXMICRO; RIGHT ATRIAL LEAD
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient SexMale
-
-