Related manufacturer reference number: 2017865-2022-45222.It was reported that the patient presented for a pacemaker change procedure.During the procedure, it was noted that the right ventricular (rv) lead was difficult to be removed from the pacemaker header, and subsequently a fluid contamination was seen between rv lead tip and pacemaker connector port.The header was cut, and the pacemaker was successfully explanted and replaced.The rv lead was reconnected to the new pacemaker.The patient was stable in condition.
|
The reported event of unable to remove right ventricular lead could not be confirmed.The device was received from the field with right ventricular lead already removed from the pacemaker.Visual inspection noticed the device header severely damaged which prevented from performing additional testing.Further visual inspection noted some dried blood which is normal, but no contamination or foreign material detected.Review of the device image indicates the device voltage is at elective replacement level.Longevity assessment was performed, based on field settings and the device exceeded projected longevity indicating normal battery depletion.
|