ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT XL; VASCULAR STENT Back to Search Results
Catalog Number EX061203C
Device Problems Fracture (1260); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, images were provided for review.The investigation of the reported event is currently underway.Medical device expiration date: 03/2023.
 
Event Description
It was reported that during a stent placement procedure, the stent allegedly had fractured.There was no reported patient injury.
 
Event Description
It was reported that during a stent placement procedure, the stent allegedly had fractured.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the catheter sample is not available for evaluation.Provided images confirm stent twist.Strut fracture cannot be confirmed because single struts are not visible.The vessel was not tortuous/ calcified, and the lesion was pre and post dilated.The user did not experience difficulty during deployment, and an irregular strut pattern was not identified upon deployment.Based on the investigation of the provided information, the investigation is closed as confirmed for stent twisting.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment; in particular the instructions for use state: 'confirm that the introducer sheath is secure and will not move during deployment.To ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.Do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.Initiate stent deployment by rotating the thumbwheel in the direction of the arrows, while holding the handle in a fixed position.' in regards to pta the instructions for use state: 'predilation of the lesion should be performed using standard techniques.', and 'post stent expansion with a pta catheter is recommended.' h10: d4 (expiration date: 03/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent xl vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent xl vascular stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the catheter sample is not available for evaluation.Provided images confirm stent twist.Strut fracture cannot be confirmed because single struts are not visible.The vessel was not tortuous/ calcified, and the lesion was pre and post dilated.The user did not experience difficulty during deployment, and an irregular strut pattern was not identified upon deployment.Based on the investigation of the provided information, the investigation is closed as confirmed for stent twisting.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment; in particular the instructions for use state: 'confirm that the introducer sheath is secure and will not move during deployment.(¿) to ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.(¿) do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.Initiate stent deployment by rotating the thumbwheel in the direction of the arrows, while holding the handle in a fixed position.' in regards to pta the instructions for use state: 'predilation of the lesion should be performed using standard techniques.', and 'post stent expansion with a pta catheter is recommended.' h10: d4 (expiration date: 03/2023), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that approximately ten days post stent placement procedure, the stent allegedly had fractured.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT XL
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15787342
MDR Text Key307599368
Report Number9681442-2022-00328
Device Sequence Number1
Product Code NIP
UDI-Device Identifier04049519001463
UDI-Public(01)04049519001463
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEX061203C
Device Lot NumberANFQ3338
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received11/14/2022
Supplement Dates Manufacturer Received11/28/2022
01/12/2023
Supplement Dates FDA Received12/14/2022
01/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight67 KG
-
-