• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TCSE-DF
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Heart Block (4444)
Event Date 11/07/2022
Event Type  Injury  
Event Description
During a premature ventricular contraction (pvc) ablation procedure, heart block occurred.Pvc mapping was done in the right ventricle, then rf application was done on the septal part of right ventricular outflow tract (rvot).Because of no efficiency, the physician decided to map with the hd grid catheter in the left ventricular outflow tract lvot.During the left mapping, a left bundle branch block appeared.Then, because of no region of interest on the left side, the user returned to the right ventricle.During a new rf application in the rvot, a complete av block occurred.No his signal was noted on the ablation catheter (distance 30 mm between the his and rf site).A temporary ventricular pacing device was placed.There was no alleged deficiency with any abbott device.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis; however, the log files from the tactisys quartz system were returned.The log file analysis concluded that the tacticath catheter performed as intended.The optical fibers met specifications, the recorded temperatures indicated cooling during rf ablation, and contact force measurements were displayed throughout the duration of the log files.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported heart block remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15793755
MDR Text Key303623797
Report Number3008452825-2022-00677
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027641
UDI-Public05415067027641
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-TCSE-DF
Device Catalogue NumberA-TCSE-DF
Device Lot Number8558452
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-