• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ NEEDLE-FREE CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ NEEDLE-FREE CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 2000E CHINA
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2022
Event Type  malfunction  
Manufacturer Narrative
2000e china 510k: this is an international code - the model#/catalog# identified is a carefusion product which is same or similar to a device that is approved for sale domestically.The domestic similar list number is 2000e.The 510k number provided is for the domestic similar product: k061285.Initial reporter phone #: (b)(6).Device expiration date: unknown.Device manufacture date: unknown.Investigation summary: a 2000e china product was not available for investigation; and the lot number was not provided in this instance.The feedback provided by the customer indicates a complete occlusion was detected during use of the smartsite device; however, no further information was available to assist the investigation.The details of this feedback were forwarded to the manufacturing site for investigation.The lot number was not available and therefore it is not possible to perform a review of the production documentation for this particular product.The root cause of the customer¿s experience could not be determined as the sample was not available for investigation.In this instance, without a sample, it is not possible to determine whether a manufacturing defect could have caused or contributed to the customer¿s experience.Please note, previous complaints for occlusions and flow restrictions have been related to features on the surface of the male luer of the connecting products.These features include flash or a raised edge to the tip of the male luer which have previously been shown to intermittently lead to restricted flow due to them pinching the blue piston of the smartsite and not allowing it to open.This can sometimes be resolved by disconnecting and reattaching the same luer connection which may reposition the luer against the piston and improve the flow, or alternatively by changing the connecting male luer (appendix 1 & 2).A review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the smartsite component in the past 12 months.
 
Event Description
It was reported that the bd smartsite¿ needle-free connector could not flow successfully.The following information was provided by the initial reporter, translated from chinese to english: the patient was admitted to the hospital with the diagnosis of "acute pancreatitis".At 10:45 on (b)(6) 2022, when the central venous tube was indwelled and connected to the needle-free sealed infusion connector, it was found that the liquid could not flow successfully, and then another needle-free sealed infusion connector was replaced.Normal use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SMARTSITE¿ NEEDLE-FREE CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15794687
MDR Text Key307591456
Report Number9616066-2022-01722
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K061285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2000E CHINA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-