Catalog Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
|
|
Event Description
|
C2 failed initial system check out - this device had been stored at morgan stanley children's hospital and was being used for training.
|
|
Manufacturer Narrative
|
Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found one failure within the data file for low 9v battery output.Visual inspection of external components found no abnormalities.Visual inspection of internal components found no abnormalities.Companion 2 driver failed incoming inspection check during the final assembly functional testing.An additional test was performed after installing a new 9v battery.Driver passed all functional testing for acceptance at inspection.Failure investigation for this complaint confirmed the reported issue.The complaint was replicated during testing; the root cause was determined to be a drained 9v battery.Failure investigation identified no other test failures or damage that could have contributed to the complaint.The companion 2 driver was not in patient use at the time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
|
Event Description
|
C2 failed initial system check out - this device had been stored at morgan stanley children's hospital and was being used for training.
|
|
Search Alerts/Recalls
|