SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problems Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2022
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event Description
C2 failed initial system check out - this device had been stored at morgan stanley children's hospital and was being used for training.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found one failure within the data file for low 9v battery output.Visual inspection of external components found no abnormalities.Visual inspection of internal components found no abnormalities.Companion 2 driver failed incoming inspection check during the final assembly functional testing.An additional test was performed after installing a new 9v battery.Driver passed all functional testing for acceptance at inspection.Failure investigation for this complaint confirmed the reported issue.The complaint was replicated during testing; the root cause was determined to be a drained 9v battery.Failure investigation identified no other test failures or damage that could have contributed to the complaint.The companion 2 driver was not in patient use at the time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event Description
C2 failed initial system check out - this device had been stored at morgan stanley children's hospital and was being used for training.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key15802148
MDR Text Key304966970
Report Number3003761017-2022-00130
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2022
Initial Date FDA Received11/15/2022
Supplement Dates Manufacturer Received11/11/2022
Supplement Dates FDA Received10/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-