SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problems Failure to Deliver Energy (1211); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2022
Event Type  malfunction  
Event Description
Customer reported tempus ls device pacer was not working.A user report was received related to a reported product problem which is currently being investigated.At the time of reporting, the device has not yet been returned for investigation.Further updates will be provided when the device is received, and the investigation is in progress.
 
Manufacturer Narrative
Philips received a complaint on the tempus ls indicating that device pacer is not working.Through the good faith effort information received that the patient was not involved.The investigations involved a "good faith effort" to obtain information and the device.Despite all the efforts, schiller manufacturer was not able to obtain them.Therefore, no investigation could be done.However, based on the failure description, this issue is most likely to be an error 26 as per schiller.This error and the exact cause is under observation at schiller in sagqi-137 (capa-0029) - tempus ls: pacer issues.
 
Event Description
Customer reported tempus ls device pacer was not working.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
rekha rv
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key15803995
MDR Text Key303722273
Report Number3003832357-2022-00051
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public7613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/19/2022
Initial Date FDA Received11/16/2022
Supplement Dates Manufacturer Received10/19/2022
Supplement Dates FDA Received02/15/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-