The hillrom technician found the right and left foot control assy needed to be replaced.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Head section motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Replace in the event of a malfunction.Make sure all functions on the caregiver control work correctly.Repair or replace the side rails as necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in (b)(6) 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the right and left¿foot control assy to resolve the reported event.Based on this information, no further action is required.
|