Catalog Number 1C8581
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the unspecified quantity of interlink system y-type solution sets would not disconnect to the patient¿s hickman (non-baxter).Several attempts were made to disconnect the set; however, the set would not disconnect and resulted in the tubing breaking off and stuck inside the hickman.This issue occurred during patient use of the device.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed the luer assembly was broken.Pressure testing was performed which revealed a leak in the affected area.Additional pull test was performed and no separation was observed.Due to the nature of the returned sample no additional testing could be performed.The reported condition was verified.The cause of the broken luer was due to the force applied by the user.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SOLUTION ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15808857
MDR Text Key307276015
Report Number1416980-2022-06283
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412014661
UDI-Public(01)00085412014661
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1C8581
Device Lot NumberDR22D26086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received12/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CART INFUSION; HICKMAN
-
-