MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 SJ4 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 SJ4 US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1231-40Q
Device Problems Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Insufficient Information (4580)
Event Date 09/29/2022
Event Type  Injury  
Event Description
It was reported that the patient presented for replacement of the implantable cardioverter defibrillator.It was noted that the patient had a history of atrial fibrillation with a rapid ventricular response for which inappropriate shocks were delivered.The device was successfully explanted and replaced with a dual-chamber generator.There were no patient consequences.
 
Manufacturer Narrative
The reported event of inappropriate therapy could not be confirmed.Interrogation of the device revealed the device was at elective replacement indicator (eri) when received.A longevity calculation was performed and found the battery depletion was normal based on the device usage.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY VR, U1.6 SJ4 US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15811457
MDR Text Key303792268
Report Number2017865-2022-45662
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734504348
UDI-Public05414734504348
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2012
Device Model NumberCD1231-40Q
Device Catalogue NumberCD1231-40Q
Device Lot Number3172601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA.
Patient Outcome(s) Other; Required Intervention;
Patient Age75 YR
Patient SexFemale
Patient Weight51 KG
-
-