• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE SECURI-T; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE SECURI-T; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number M00509161
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2022
Event Type  Injury  
Manufacturer Narrative
(device codes): (b)(4).
 
Event Description
It was reported to boston scientific corporation that an endovive securi-t replacement bolster was used during a percutaneous endoscopic gastrostomy replacement procedure six months ago.During the replacement procedure, on (b)(6) 2022, the internal bolster detached.An endoscopic examination was immediately performed; however, bolster was not found.A new endovive securi-t replacement bolster was placed.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6 (device codes): device code a0501 captures the reportable event of internal bolster detached.Block h10: the returned endovive securi-t replacement bolster was analyzed.Upon visual assessment, it was observed that the bolster of the silicon tube was broken.In addition, part of the bolster was missing, and silicone tube contains remnants of use.Therefore the complaint is confirmed for bolster detached.Based on the condition of the returned device, engineers determined that based upon the adhesion traces found in the silicone tube as well as the detachment and damage to the bolster it suggests that the device was subjected to excessive force; perhaps the manipulation, or patient's anatomical conditions could have contributed to this event.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on the event which led to the reported event.
 
Event Description
It was reported to boston scientific corporation that an endovive securi-t replacement bolster was used during a percutaneous endoscopic gastrostomy replacement procedure six months ago.During the replacement procedure, on (b)(6) 2022, the internal bolster detached.An endoscopic examination was immediately performed; however, retrieval of detached portion was unsuccessful.A new endovive securi-t replacement bolster was placed.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE SECURI-T
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15813605
MDR Text Key303800139
Report Number3005099803-2022-06627
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00509161
Device Catalogue Number0916
Device Lot Number0029548529
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2022
Initial Date FDA Received11/17/2022
Supplement Dates Manufacturer Received11/28/2022
Supplement Dates FDA Received12/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-