BOSTON SCIENTIFIC CORPORATION ENDOVIVE SECURI-T; TUBES, GASTROINTESTINAL (AND ACCESSORIES)
|
Back to Search Results |
|
Model Number M00509161 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/24/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
(device codes): (b)(4).
|
|
Event Description
|
It was reported to boston scientific corporation that an endovive securi-t replacement bolster was used during a percutaneous endoscopic gastrostomy replacement procedure six months ago.During the replacement procedure, on (b)(6) 2022, the internal bolster detached.An endoscopic examination was immediately performed; however, bolster was not found.A new endovive securi-t replacement bolster was placed.There were no patient complications reported as a result of this event.
|
|
Manufacturer Narrative
|
Block h6 (device codes): device code a0501 captures the reportable event of internal bolster detached.Block h10: the returned endovive securi-t replacement bolster was analyzed.Upon visual assessment, it was observed that the bolster of the silicon tube was broken.In addition, part of the bolster was missing, and silicone tube contains remnants of use.Therefore the complaint is confirmed for bolster detached.Based on the condition of the returned device, engineers determined that based upon the adhesion traces found in the silicone tube as well as the detachment and damage to the bolster it suggests that the device was subjected to excessive force; perhaps the manipulation, or patient's anatomical conditions could have contributed to this event.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on the event which led to the reported event.
|
|
Event Description
|
It was reported to boston scientific corporation that an endovive securi-t replacement bolster was used during a percutaneous endoscopic gastrostomy replacement procedure six months ago.During the replacement procedure, on (b)(6) 2022, the internal bolster detached.An endoscopic examination was immediately performed; however, retrieval of detached portion was unsuccessful.A new endovive securi-t replacement bolster was placed.There were no patient complications reported as a result of this event.
|
|