• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Event Description
It was reported that before an unknown procedure, the qr scanner could not read the devices lot number qr code.No patient consequence.
 
Manufacturer Narrative
(b)(4).Date of event: unknown, assumed first day of month that complaint was reported.Batch # x9594p.An analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.
 
Manufacturer Narrative
(b)(4).Date sent: 12/15/2022.Photo analysis this is an analysis of a photo submitted to ethicon for evaluation.During the visual analysis the following was observed: the photo shows a device product code el5ml batch number x9594p in its original packaging, the qr code cannot be read by the scanner.A clear analysis of the qr code could not be performed as the photo becomes distorted when magnified.Based on the photo review, the event describe is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15815634
MDR Text Key307774506
Report Number3005075853-2022-08030
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX9594P
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/01/2022
Initial Date FDA Received11/17/2022
Supplement Dates Manufacturer Received11/19/2022
Supplement Dates FDA Received12/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-