Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Event Description
It was reported that during the pretest it was discovered the device was not running smoothly.No adverse events were reported as a result of this malfunction.No further information is available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
No further information is available at this time.
 
Manufacturer Narrative
This event is recorded with zimmer biomet under: (b)(4).This medwatch is being filed as a final report based on information discovered during the device evaluation.Review of the most recent repair record determined the motor ran within specifications; however, it was running erratically.The motor, plug harness assembly and switch were replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15819429
MDR Text Key303902472
Report Number0001526350-2022-01123
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number61936628
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-