(b)(4).Batch # x94f6g.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no damage in the external components.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and the clips were not fed into the jaws.In order to evaluate the condition of the internal components the device was disassembled; upon disassembling the feeder shoe tab was found to be damaged causing the feeding issues.In addition, 12 clips were found inside the clip track.No conclusion could be reached regarding what may have caused the reported incident.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|