Catalog Number V041183000000 |
Device Problems
Detachment of Device or Device Component (2907); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it is reported that vdw.Connect drive inkl.6:1 ca won't hold files.No injury occurred.
|
|
Manufacturer Narrative
|
Investigation summary: nc073950: handpiece 10396 (does not turn on) defect, replaced with 11524.Contra angle 05584 (head drive worn) defect, replaced with 15318.Charging unit not provided for investigation.Defective component/part: contra angle (head drive worn) + handpiece (does not turn on) all complaints are monitored through the monthly product surveillance committee and a corrective action could be determined by the committee.
|
|