The reported event of no pacing could not be confirmed.Final analysis found that as received, the device battery voltage was above elective replacement indicator (eri) level.The device was then programed to nominal settings for the remainder of the analysis.The results of all electrical test performed, including mechanical stress testing, output test and battery assessment indicate normal device characteristics.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
|