The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging difficulty breathing/short of breath, sleeping problem and oxygen levels has been low.There was no report of patient serious harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an alleging difficulty breathing/short of breath, sleeping problem and oxygen levels has been low of an issue related to a cpap device's sound abatement foam.There was no report of serious or permanent harm or injury.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.The manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation.Updated the section d8, d9, h3, and h6.
|