Updated fields: d4, d9, g4, g6, h2, h3, h4, h6, h10 the mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: evaluation found no device deficiencies that would have contributed to the reported complaint of "slippage." they were unable to duplicate slippage.However, unit needed basic preventive maintenance from routine use.Root cause: the complaint is not confirmed via inspection of the unit.Probable root cause of the reported complaint is improper or suboptimal placement of the mayfield system.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|