SMITHS MEDICAL ASD, INC. CADD HIGH VOL ADMINISTRATION SET W/FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7346-24
Device Problems Fluid/Blood Leak (1250); Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2022
Event Type  malfunction  
Event Description
It was reported that priming the line unsuccessful.It would normally require less then 10 milliliters to prime a line, but this time the pump continued to prime until approximately 20 milliliters.The line was unable to deliver medication as the filter on line was noted to be leaking.No serious harm was reported.There was no patient injury reported.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.Operator of device is unknown.No information has been provided to date.
 
Manufacturer Narrative
No product or photographic evidence were provided to aid in this investigation.As a result, a complaint investigation / product evaluation and problem confirmation cannot be performed.Lacking any additional evidence in the complaint file attachment, and details in the complaint description, a probable cause could not be determined.Therefore, this complaint has been closed as unconfirmed.If the product is returned, this complaint will be reopened for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).
 
Manufacturer Narrative
H6: health effects, clinical signs and symptoms or conditions updated, corrected data: h6: evaluation codes conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD HIGH VOL ADMINISTRATION SET W/FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
14th floor, north tower
minneapolis, MN 55442
MDR Report Key15830413
MDR Text Key307515521
Report Number3012307300-2022-27363
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Device Lot Number4248635
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2022
Initial Date FDA Received11/19/2022
Supplement Dates Manufacturer Received03/17/2023
10/16/2023
Supplement Dates FDA Received06/22/2023
10/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-