Catalog Number 03666794001 |
Device Problems
Image Display Error/Artifact (1304); Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/02/2022 |
Event Type
malfunction
|
Event Description
|
There was an allegation of a display issue with the coaguchek xs meter.After replacing the batteries, the customer alleged the display was missing segments and the "00" in date format looked like an "l7".When performing a display check, the customer stated one segment was missing from an "8" in the results.There was no allegation of a result misinterpretation.
|
|
Manufacturer Narrative
|
The meter was requested to be returned.The investigation is ongoing.
|
|
Manufacturer Narrative
|
The meter was returned for investigation.The device could not be powered on.The battery contacts and the circuit board are contaminated by liquid.The contamination effects the battery contacts on the circuit board, which are interrupted by the contamination.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 have been updated.
|
|