Catalog Number 03666794001
Device Problems Image Display Error/Artifact (1304); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2022
Event Type  malfunction  
Event Description
There was an allegation of a display issue with the coaguchek xs meter.After replacing the batteries, the customer alleged the display was missing segments and the "00" in date format looked like an "l7".When performing a display check, the customer stated one segment was missing from an "8" in the results.There was no allegation of a result misinterpretation.
 
Manufacturer Narrative
The meter was requested to be returned.The investigation is ongoing.
 
Manufacturer Narrative
The meter was returned for investigation.The device could not be powered on.The battery contacts and the circuit board are contaminated by liquid.The contamination effects the battery contacts on the circuit board, which are interrupted by the contamination.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15833466
MDR Text Key307723178
Report Number1823260-2022-03697
Device Sequence Number1
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2022
Initial Date FDA Received11/21/2022
Supplement Dates Manufacturer Received12/09/2022
Supplement Dates FDA Received12/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-