• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number EVD35-06-150-120
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2022
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A physician was attempting to implant an everflex self-expanding stent with the entrust delivery system for the treatment of a 130mm calcified lesion in the mid superficial femoral artery.The artery had a diameter of 6mm with moderate tortuosity and severe calcification.A 6fr non-mdt sheath was used and a 7mm spider fx was used.The device was prepped as per the ifu with no issues identified.The lock pin was removed before starting delivery at the lesion.It was reported the stent partially deployed. the strain relief was cut with a scalpel.The white handle was then separated in half exposing the thumbwheel and thread.A pair of hemostats was used to pull on the inner sheath while watching the stent deploy under flouroscopy.The stent was deployed successfully.It was reported the spider wire kinked.No patient injury was reported.
 
Manufacturer Narrative
Image analysis the customer returned one image for evaluation image 1: the image depicts the everflex entrust handle in two halves, with the thumbwheel out of the devices handle and the pullwire is observed to be broken.In the picture the strain relief is off the device and beside the broken handle, the strain relief shows a size of 6 x 150mm, which correlates with the size in the event description.There are other various ancillary devices / components within the picture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15833651
MDR Text Key307559965
Report Number2183870-2022-00393
Device Sequence Number1
Product Code NIP
UDI-Device Identifier00821684051399
UDI-Public00821684051399
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEVD35-06-150-120
Device Catalogue NumberEVD35-06-150-120
Device Lot NumberB419412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2022
Initial Date FDA Received11/21/2022
Supplement Dates Manufacturer Received11/03/2022
Supplement Dates FDA Received12/04/2022
Date Device Manufactured07/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1644-2022
Patient Sequence Number1
Patient SexMale
-
-