• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3222
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2022
Event Type  malfunction  
Event Description
It was reported that during an unrelated procedure, the hex wrench did not disengage from the set screw resulting in the inability to re-insert the set screw into the header of the device.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of left ventricle-setscrew stuck to the wrench was confirmed.Analysis of the device revealed that the user of the torque wrench did not align the wrench tip to go into the inset of the setscrew.The wrench was inserted with the corners of the wrench tip going into, being wedged and forced into the inset walls.This resulted in the wrench being stuck in the inset of the setscrew.The setscrew was then pulled out of the device through the septum attached to the wrench tip.This is related to the user's incorrect use of the wrench.A non-abbott torque wrench was also used which is not recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15838232
MDR Text Key306828373
Report Number2017865-2022-46145
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507905
UDI-Public05414734507905
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model NumberPM3222
Device Catalogue NumberPM3222
Device Lot NumberP000139480
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-