Model Number 518-019
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Perforation (2513); Pericardial Effusion (3271)
Event Date 11/04/2022
Event Type  Injury  
Manufacturer Narrative
The device was discarded, thus no investigation could be completed.Cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event Description
A lead extraction procedure commenced to remove a right ventricular (rv) lead due to fracture/non-function.A spectranetics lead locking device (lld) was inserted into the lead to provide traction.A spectranetics 14f glidelight laser sheath was used, encountering stalled progress about halfway down the superior vena cava (svc).The glidelight's teflon outer sheath was then added, but progress stalled in the same location.The physician upsized to a 16f glidelight without an outer sheath in order to advance past the area, and was able to progress to where the tip of the glidelight reached the ra/rv junction.The patient's blood pressure dropped; the rv lead came free and was extracted.A pericardial effusion was detected.Rescue efforts began immediately, including rescue balloon, pericardiocentesis, bypass and sternotomy.A 2-3 cm posterior ra perforation was discovered and repaired.The patient survived the procedure.The physician believed that traction or lasing caused the perforation when the lead was pulled away from the binding site; however, significant tissue was present on the lead when it was extracted.This indicated that traction had caused or contributed to the perforation.This report captures the lld providing traction to the rv lead when the perforation occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15838611
MDR Text Key304087507
Report Number1721279-2022-00211
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023010
UDI-Public(01)00813132023010(17)231217(10)FLC21M15A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990713
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/17/2023
Device Model Number518-019
Device Catalogue Number518-019
Device Lot NumberFLC21M15A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/04/2022
Initial Date FDA Received11/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC 6935M RV ICD LEAD; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH; SPECTRANETICS 16F GLIDELIGHT LASER SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM
Patient Outcome(s) Life Threatening; Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight101 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-