• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FREEZOR XTRA CARDIAC CRYOABLATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FREEZOR XTRA CARDIAC CRYOABLATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 217F3
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2022
Event Type  malfunction  
Event Description
It was reported that during a cryo ablation procedure, the ablations were ineffective.Due to the patient's anatomy, it was difficult to maneuver the catheter without perforating the artery.  the catheter length also contributed to the decision to switch the case to radiofrequency and the physician wanted to perform left side bypass ablation. the case was completed with radiofrequency. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the 217f3 electrophysiology (ep) catheter with lot 13254 was returned and analyzed.During external visual inspection of the shaft segment, a kink was observed on the shaft.The catheter smart chip data was reviewed.Data indicated that the catheter was used for 24 applications on the reported event date.During system performance testing with the test console, the console terminated the application and generated system notice 50005 (indicating that the safety system detected fluid in the catheter and stopped the injection).During pressure testing and inspection of the shaft segment, a shaft breach was observed at shaft.In conclusion, the ep catheter failed the returned product inspection due to a shaft kink and breach.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEZOR XTRA CARDIAC CRYOABLATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15839154
MDR Text Key307831085
Report Number3002648230-2022-00451
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/22/2022
Device Model Number217F3
Device Catalogue Number217F3
Device Lot Number13254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2022
Initial Date Manufacturer Received 10/26/2022
Initial Date FDA Received11/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
Patient SexFemale
-
-