• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G154
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bacterial Infection (1735); Inflammation (1932); Swelling/ Edema (4577)
Event Date 10/31/2022
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.Reportedly, the patient developed redness and swelling.The patient was treated with intravenous antibiotics.There were no additional adverse patient effects reported.The crt-d was explanted.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against the device was not confirmed.The device was analyzed, passed all the baseline tests and exhibited normal device functions.Analysis of the returned product is not able to provide relevant information for infection-related allegations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15840189
MDR Text Key304081111
Report Number2124215-2022-48358
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534652
UDI-Public00802526534652
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/13/2024
Device Model NumberG154
Device Catalogue NumberG154
Device Lot Number106434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2022
Initial Date FDA Received11/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexMale
-
-