• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT Back to Search Results
Catalog Number 5F060401C
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2022
Event Type  malfunction  
Event Description
It was reported that during a stent placement procedure in the external iliac artery via a contralateral approach, the tip of the catheter allegedly confirmed to be detached in the distal part of left common femoral artery.It was further reported that the tip of the catheter was removed from left common femoral artery.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent 5f vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent products are identified in common device name and pma/510k.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.Expiry date: 06/2023.
 
Event Description
It was reported that during a stent placement procedure in the external iliac artery via a contralateral approach, the tip of the catheter allegedly confirmed to be detached in the distal part of left common femoral artery.It was further reported that the tip of the catheter was removed from left common femoral artery.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent 5f vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent 5f vascular stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the physical sample was returned for evaluation.The returned sample was found in used condition without stent.The inner catheter cardan tube was found broken twice which leads to confirmed result for inner catheter break.It was known that a 6f introducer sheath was used for access.Based on the information available the investigation is closed with confirmed result for inner catheter cardan break.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the instructions for use states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' under materials required the instructions for use states: '5f (1.67 mm) or larger introducer sheath (¿); 0.014-inch (0.36 mm) - 0.035-inch (0.89 mm) diameter guidewire'.Regarding insertion and removal difficulty of the delivery system the instructions for use states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.H10: d4 (expiry date: 06/2023), g3, h6 (device).H11: h6 (result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15841637
MDR Text Key307984207
Report Number9681442-2022-00336
Device Sequence Number1
Product Code NIP
UDI-Device Identifier00801741119972
UDI-Public(01)00801741119972
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5F060401C
Device Lot NumberANET2465
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-