MAUDE Adverse Event Report: ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE LTD FREESTYLE LIBRE 3; CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 72081-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Erythema (1840); Burning Sensation (2146); Skin Infection (4544)
Event Date 11/08/2022
Event Type  Injury  
Event Description
An adverse skin reaction was reported with wear of the adc device.Customer experienced symptoms described as "puncture site has started to burn, red circle after sensor removal" and had contact with a healthcare professional for unspecified treatment.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.If the product is returned, a physical investigation will be performed and a follow-up report submitted.N/a was selected for pma/510k as it is unknown if the user was using android, ios, or a reader with the fs libre 3 sensor.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15844433
MDR Text Key304114574
Report Number2954323-2022-42773
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/28/2023
Device Model Number72081-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/11/2022
Initial Date FDA Received11/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-