ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Product Quality Problem (1506)
Patient Problem Electric Shock (2554)
Event Date 11/03/2022
Event Type  malfunction  
Event Description
Customer reports an electric shock while wearing their adc device.There was no report of death, serious injury, or mistreatment associated with this event.
 
Manufacturer Narrative
The reported product has been returned and investigation is pending at this time.A follow-up report will be filed once additional information is obtained.The device manufacturer date for the reported sensor is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection was performed on the returned sensor and no issues were observed.The sensor plug was properly seated in the mount.Visual inspection of the plug assembly and no failures were observed.Visual inspection was performed on the de-cased sensor's pcba (printed circuit board assembly) and no issues were observed.Extended investigation was also performed on the returned sensor.Visual inspection was performed on the returned sensor and no abnormalities were found.The battery was measured at 1.58v and voltage was within specification.The sensor was reprogrammed and successful accuracy testing was performed.A thermal camera was used to monitor the sensor while in glucose detection mode, no hot spots on the pcba were observed.The customer complained they felt an electrical shock.The puck's battery is rated at 1.6v which is not enough voltage to produce a shock and there was no evidence of a device failure.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reports an electric shock while wearing their adc device.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15845001
MDR Text Key307309531
Report Number2954323-2022-42778
Device Sequence Number1
Product Code QLG
UDI-Device Identifier00357599800000
UDI-Public00357599800000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2023
Device Model Number71992-01
Device Catalogue Number71992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2022
Initial Date Manufacturer Received 11/03/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received12/29/2022
Supplement Dates FDA Received01/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-