• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3222
Device Problems Loose or Intermittent Connection (1371); Difficult to Remove (1528); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for routine device change.During the procedure, the setscrew wrench was stuck, and the lead was unable to be removed from the device header.The physician alleged that the right and left ventricular ports were stripped.Multiple tools were used to break the device header and free the lead.The device was replaced.The patient was stable.
 
Manufacturer Narrative
The device was above elective replacement indicator (eri) upon receipt.A visual inspection found the header broken off from the device can and it was not returned for analysis.The broken header was consistent with damage from the explant procedure.Further analysis performed showed normal device characteristics.A longevity assessment was performed, and the device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15847417
MDR Text Key304544052
Report Number2017865-2022-46276
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507905
UDI-Public05414734507905
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberPM3222
Device Catalogue NumberPM3222
Device Lot NumberA000053511
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LEFT VENTRICULAR EPICARDIAL LEAD; RIGHT ATRIAL LEAD; RIGHT VENTRICULAR EPICARDIAL LEAD
Patient SexMale
-
-