• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279401200
Device Problem Continuous Firing (1123)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there was continuous activation.
 
Manufacturer Narrative
Alleged failure: serfas activates itself.Without pressing the button.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be probe short due to fluid ingress.A broken/damaged bond of the polyimide and suction tubing causes fluid ingress to the pcb which can lead to a short circuit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was continuous activation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15848072
MDR Text Key306918641
Report Number0002936485-2022-00695
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0279401200
Device Catalogue Number0279401200
Device Lot Number22140AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received10/27/2022
Supplement Dates FDA Received01/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-