Catalog Number 306594
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2022
Event Type  malfunction  
Event Description
It was reported that the end of the bd posiflush¿ syringe plunger rod was found damaged during use.The following information was provided by the initial reporter, translated from chinese: "on february 26, 2022, the patient's infusion was finished, and when the nurse sealed the catheter with indwelling needle for the patient, she found that the tail of the core rod of the 5ml flush was incomplete, which could not be injected.After communication with the nurse, the patient expressed understanding and replaced the 5ml flush with the sealing tube.".
 
Manufacturer Narrative
Initial reporter e-mail: unknown.Investigation summary: a device history record review was completed by our quality engineer team for provided material number 306594 and lot number 1140604.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15849097
MDR Text Key307668192
Report Number1911916-2022-00670
Device Sequence Number1
Product Code NGT
UDI-Device Identifier00382903065943
UDI-Public(01)00382903065943
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number306594
Device Lot Number1140604
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-