• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE PROGRESSA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE PROGRESSA FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P7500A000295
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2022
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed had no audible brake not set alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
 
Manufacturer Narrative
The hillrom technician found the external alarm needed to be replaced.Per the progressa user manual: the brake not set is an audible and visual alarm.The alarm will sound, and a message will show on the gci, when the bed is connected to an ac power source and the brake is not set.When the bed is connected to ac power and the brakes are not set, an alarm sounds and a message shows on the gci.When ac power is removed, the alarm will stop and the gci will turn off.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the external alarm to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRESSA FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
maritza valencia
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key15857232
MDR Text Key305629305
Report Number1824206-2022-00510
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761000100
UDI-Public010088776100010011170605
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7500A000295
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-